MicroDigital

Diamond - Introduction

Diamond
Fully Automatic Immuno Analyzer
Diamond – Fully Automatic Immuno-Analyzer takes care of sample and reagent handling to well washing and optical measurement with a touch of a button.

Diamond is a convenient and accurate fully automated system specialized for CLIA(chemiluminescence immunoassay) and ELISA(enzyme-linked immunosorbent assay)
Feature
- Fully automated immunoassay system (luminescence, absorbance measurement)
- Open platform
- Compatible with conventional 96-well plates
- Quick and accurate control
- Minimized inter-well cross contamination through the use of disposable tips
- Accurate liquid handling
Specification
General
Sample Capacity 96 Patient Sample Positions
Reagent Capacity 9 Custom Reagent Positions
Test Tube Standard 13, 16 mm diameter (Customizable)
No. of Microplate 2
Well-tracking Independent well tracking & timing
Barcode Reader Optional
System Architecture Universal Open Platform (Qualitative and Quantitative)
Sample Tray Easy loading and unloading of samples
Dimension 650 x 830 x 650 mm
Power 100 – 240 V, 50 – 60 Hz
ELISA Reader Specification
Dynamic Range 0 – 3.0 OD
Photometer 450, 630 nm (Optional: 405, 490, 550 nm)
Filter Slots 6
Precision <1% CV (2.0 OD)
Accuracy ±0.005 OD or 2.5%
CLIA Reader Specification
Dynamic Range Up to 7-log of dynamic range
Limit of Detection 1 x 10-18 moles of ATP
Crosstalk 3.0 x 10-5
Washer
Manifold Configuration 8 Channels
Dispense-volume Range 100 – 300 µl
Washing Tank 2 External Washing Tanks and 2 Waste Tanks
Wash Cycle 1 - ∞ (Customizable)
Residual Volume <1 µl
Incubator
Temperature Range Up to 42℃
Temperature Uniformity ±0.5℃
Shaking Amplitude: 1 mm, Speed: Up to 800 RPM
Temperature Monitoring Yes
Pipetting
Aspiration Volume >5 µl
Tip Disposable
Serial Dilution Yes
Replicates Yes
Precision <0.8% CV at 100 µl
Process Security
Liquid-level Detection Yes (Optional) (Must use disposable tips from MicroDigital)
Level-sensor System Pressure Differential
Dispense-anomaly Detection Yes (Optional)
Tip Detection Yes (Optional)
Well-fill Verification Yes
Alarms Warning Message on PC Software
Software
Operating Software 3 Types (Developer, User, and Hospital Mode)
Work Protocols (Assays) Unlimited
Data Processing Quantitative and Qualitative
Access Level Control Yes (Master, Manager, and Operator)
O/S Windows XP, 7, and 10
Models
Diamond E System for ELISA
Diamond C System for CLIA
Diamond M System for ELISA & CLIA